.LP
|
|
.af PN i
|
|
.ds LH MIT-BIH A\s-2RRHYTHMIA\s+2 D\s-2ATABASE\s+2
|
|
.ds RH F\s-2OREWORD\s+2
|
|
.bp 3
|
|
\s+2
|
|
\fB
|
|
.ce 1
|
|
FOREWORD
|
|
.PP
|
|
For a number of years our group has been investigating methods for real-time
|
|
ECG rhythm analysis. In the course of this work, we have developed an
|
|
extensive annotated digital ECG database. The database has been enormously
|
|
helpful to us in algorithm development and evaluation. The creation of this
|
|
resource required a major effort, and was funded, in part, by both government
|
|
and industry. We feel it is highly desirable to make this database available
|
|
to other academic and industrial groups, and hence have prepared it for
|
|
distribution. This catalog contains detailed descriptions of the database
|
|
tapes.
|
|
.PP
|
|
We acknowledge with gratitude the many dedicated hours of work which
|
|
went into this project on the part of cardiologists, Holter technicians,
|
|
laboratory assistants, and engineers in our laboratories at MIT and at
|
|
Beth Israel Hospital. We also acknowledge the help of our colleagues at
|
|
Washington University, St. Louis (particularly Russell Hermes) in
|
|
assuring a compatible data format. We especially wish to recognize:
|
|
.IP \(bu
|
|
\fIPaul Schluter\fR, who did the original design and implementation of the
|
|
database.
|
|
.IP \(bu
|
|
\fIScott Peterson\fR, who supervised the detailed data selection, digitization,
|
|
annotation, and editing at Beth Israel Hospital. He also contributed
|
|
substantially to the development of software needed for using the database in
|
|
our evaluation.
|
|
.IP \(bu
|
|
\fIGeorge Moody\fR, who converted the database to a format compatible with that
|
|
used by the AHA database, and who contributed in a major way to the directory.
|
|
.IP \(bu
|
|
\fILarry Siegal\fR, who contributed to development of the waveform editor
|
|
system.
|
|
.IP \(bu
|
|
\fICheryl Jackson\fR, who was responsible for most of the detailed
|
|
transcription of cardiologist annotations, the comparison and quality control
|
|
functions, and who assembled the final manuscript for the directory.
|
|
.IP \(bu
|
|
\fIDiane Perry\fR, who as chief technician in the Arrhythmia Laboratory helped
|
|
to identify suitable data, and who helped in the annotation process.
|
|
.IP \(bu
|
|
The group of physicians who helped with the difficult task of beat-by-beat
|
|
annotation of the ECGs:
|
|
Dr. Esmerey Acarturk, Dr. John Aumiller, Dr. Sidney Blake, Dr. Alvin
|
|
Blaustein, Dr. Chester Conrad, Dr. Gary Heller, Dr. Michael Malagold,
|
|
Dr. Roger Mark, and Dr. Candice Miklozek.
|
|
|
|
.nf
|
|
.in +3i
|
|
Roger Mark
|
|
Walter Olson
|
|
|
|
Cambridge, Massachusetts
|
|
September, 1980
|
|
.in -3i
|
|
.fi
|
|
|
|
\s+2
|
|
\fB
|
|
.ce 1
|
|
NOTES ON THE SECOND EDITION
|
|
.PP
|
|
During the eight years since we first published this book, nearly one hundred
|
|
academic and industrial research groups worldwide have used the MIT-BIH
|
|
Arrhythmia Database. We thank these organizations for their support.
|
|
.PP
|
|
In addition to those listed above, we also wish to recognize the
|
|
contributions of:
|
|
.IP \(bu
|
|
\fIJoe Mietus\fR, who handled correspondence as well as production and
|
|
distribution of the database from the Beth Israel Hospital, and who analyzed
|
|
the mechanical sources of analog tape wow and flutter.
|
|
.IP \(bu
|
|
\fITed Baker\fR, who helped in the first ``port'' of the database from our
|
|
homebrew 8080-based systems.
|
|
.PP
|
|
In lieu of reprinting the first edition of this volume, we have made use of
|
|
modern printing technology to produce a far more readable and complete record
|
|
of the contents of the database.
|
|
.PP
|
|
In June, 1987, the Association for the Advancement of Medical Instrumentation
|
|
published its \fIRecommended Practice for Testing and Reporting
|
|
Performance Results of Ventricular Arrhythmia Detection Algorithms\fR
|
|
(AAMI ECAR-1987), which may be obtained from AAMI, 3330 Washington Boulevard,
|
|
Suite 400, Arlington, VA 22201.
|
|
The MIT-BIH and AHA Databases provide developers and evaluators of arrhythmia
|
|
detectors with standard test data; the AAMI Recommended Practice provides
|
|
guidelines for using these databases in a standard way, and for describing
|
|
detector performance in a manner that facilitates comparisons between
|
|
detectors.
|
|
We urge all users of our database to follow the AAMI recommendations when
|
|
preparing performance statistics for publication.
|
|
.PP
|
|
A package of C-language software for using the MIT-BIH and AHA Databases
|
|
is available from MIT. The package
|
|
includes programs for plotting ECGs with annotations, sampling rate
|
|
conversion, and beat-by-beat comparison of annotation files following the
|
|
AAMI recommended practice, as well as a variety of other useful programs.
|
|
All the programs access the database via a common library of subroutines,
|
|
which are also provided as part of the package and which may be used with
|
|
user programs.
|
|
.PP
|
|
We have made several other sets of ECG recordings available, including
|
|
specialized databases for ventricular fibrillation, atrial fibrillation,
|
|
and ST segment changes. These databases are not annotated beat-by-beat.
|
|
|
|
.nf
|
|
.in +3i
|
|
George Moody
|
|
Roger Mark
|
|
|
|
Cambridge, Massachusetts
|
|
August, 1988
|
|
.in -3i
|
|
.fi
|
|
|
|
\s+2
|
|
\fB
|
|
.ce 1
|
|
NOTES ON THE THIRD EDITION
|
|
.PP
|
|
Since the publication of the second edition, the database has been made
|
|
available in CD-ROM format. Continued strong interest in the database has made
|
|
it possible to prepare a second edition of the CD-ROM and a third edition of
|
|
this book. The CD-ROM includes the additional specialized databases mentioned
|
|
above, and the second edition of the CD-ROM also contains the software
|
|
package mentioned above.
|
|
.PP
|
|
Compatible databases of ECGs and other physiologic signals are beginning to
|
|
appear. Of particular interest to users of our database is the European ST-T
|
|
Database, which consists of ninety two-hour ECG recordings with beats, rhythms,
|
|
and signal quality annotated as in the MIT-BIH Arrhythmia Database, and with
|
|
additional annotations to indicate ST and T-wave morphology changes. For
|
|
information, write to: CNR Institute of Clinical Physiology, Computer
|
|
Laboratory, via Trieste, 41, 56100 Pisa, Italy.
|
|
.PP
|
|
I wish to thank all of those who have supported this project over these years,
|
|
especially Roger Mark, who has guided it from its inception in 1975.
|
|
|
|
.nf
|
|
.in +3i
|
|
George Moody
|
|
|
|
Cambridge, Massachusetts
|
|
July, 1992
|
|
.in -3i
|
|
.fi
|
|
.ds RH I\s-2NTRODUCTION\s+2
|
|
.bp
|
|
.SH
|
|
INTRODUCTION
|
|
.PP
|
|
This introduction describes how the database records were
|
|
obtained, and discusses the characteristics of the recorded signals.
|
|
Following these notes are annotated ``full disclosure'' plots of the entire
|
|
database. These can be useful for obtaining an overall
|
|
impression of the contents of individual records. Following the
|
|
``full disclosure'' plots are sample ECG strips. These strips
|
|
were chosen to illustrate the salient features of each record.
|
|
Next are notes on the important features of each record.
|
|
These notes also include background information on the subjects,
|
|
including their medications.
|
|
At the end of the book are tables of rhythms and annotations, which
|
|
summarize the contents of the database.
|
|
These tables can be helpful in finding a record with a specific
|
|
set of characteristics.
|
|
.SH
|
|
Selection criteria
|
|
.PP
|
|
The source of the ECGs included in the MIT-BIH Arrhythmia Database is a
|
|
set of over 4000 long-term Holter recordings that were obtained
|
|
by the Beth Israel Hospital Arrhythmia Laboratory between 1975 and 1979.
|
|
Approximately 60% of these recordings were obtained from inpatients.
|
|
The database contains 23 records
|
|
(numbered from 100 to 124 inclusive with some numbers missing)
|
|
chosen at random from this set, and 25 records
|
|
(numbered from 200 to 234 inclusive, again with some numbers missing)
|
|
selected from the same set to include a variety of rare but clinically
|
|
important phenomena that would not be well-represented
|
|
by a small random sample of Holter recordings.
|
|
Each of the 48 records is slightly over 30 minutes long.
|
|
.PP
|
|
The first group is intended to serve as a representative sample of the
|
|
variety of waveforms and artifact that an arrhythmia detector might
|
|
encounter in routine clinical use. A table of random numbers was used
|
|
to select tapes, and then to select half-hour segments of them.
|
|
Segments selected in this way were excluded only if neither of the two
|
|
ECG signals was of adequate quality for analysis by human experts.
|
|
.PP
|
|
Records in the second group were chosen to
|
|
include complex ventricular, junctional,
|
|
and supraventricular arrhythmias and conduction abnormalities. Several
|
|
of these records were selected because features of the rhythm, QRS
|
|
morphology variation, or signal quality may be expected to present
|
|
significant difficulty to arrhythmia detectors; these records have
|
|
gained considerable notoriety among database users.
|
|
.PP
|
|
The subjects were 25 men aged 32 to 89 years, and 22 women aged 23 to 89
|
|
years. (Records 201 and 202 came from the same male subject.)
|
|
.SH
|
|
ECG lead configuration
|
|
.PP
|
|
In most records, the upper signal is a modified limb lead II (MLII),
|
|
obtained by placing the electrodes on the chest. The lower signal is
|
|
usually a modified lead V1 (occasionally V2 or V5, and in one instance V4);
|
|
as for the upper signal, the electrodes are also placed on the chest.
|
|
This configuration is routinely used by the BIH Arrhythmia Laboratory.
|
|
Normal QRS complexes are usually prominent in the upper signal.
|
|
The lead axis for the lower signal may be nearly orthogonal to the mean
|
|
cardiac electrical axis, however (i.e., normal beats are usually
|
|
biphasic and may be nearly isoelectric).
|
|
Thus normal beats are frequently difficult to discern in the lower signal,
|
|
although ectopic beats will often be more prominent (see, for example, record
|
|
106).
|
|
A notable exception is record 114, for which the signals were reversed.
|
|
Since this happens occasionally in clinical practice, arrhythmia detectors
|
|
should be equipped to deal with this situation.
|
|
In records 102 and 104, it was not possible to use modified lead II because of
|
|
surgical dressings on the patients;
|
|
modified lead V5 was used for the upper signal in these records.
|
|
.SH
|
|
Analog recording and playback
|
|
.PP
|
|
The original analog recordings were made using nine Del Mar Avionics
|
|
model 445 two-channel recorders, designated \fIA\fP through \fII\fP:
|
|
.TS
|
|
center box;
|
|
c | l.
|
|
\fIRecorder Records\fP
|
|
_
|
|
A 102, 107, 111, 115, 121
|
|
B 212
|
|
C 203
|
|
D 118, 124, 217
|
|
E 101, 103, 106, 108, 112, 117, 119, 122, 209, 219, 220, 223, 233
|
|
F 104, 109, 123, 205, 207, 210, 215, 221
|
|
G 100, 105, 114, 116, 213, 214, 222, 228
|
|
H 113, 201, 202, 231
|
|
I 200, 230, 232, 234
|
|
.TE
|
|
(It is not known which recorder was used for record 208.)
|
|
.PP
|
|
During the digitization process, the analog recordings were played back
|
|
on a Del Mar Avionics model 660 unit. The analog tapes used for records
|
|
112, 115 through 124, 205, 220, 223, and 230 through 234 were played back
|
|
and digitized at twice real time; the rest were played back at real time
|
|
using a specially constructed capstan for the model 660 unit.
|
|
Skew between the two signals was found to be as great as 40
|
|
milliseconds for some of the analog recorders.
|
|
In addition to the fixed skew that results from extremely small differences
|
|
in the orientations of the tape heads on the recorder and the playback unit,
|
|
microscopic vertical wobbling of the tape, either during recording or playback,
|
|
introduces a variable skew, which may be comparable in magnitude to the fixed
|
|
skew.
|
|
This problem (which also
|
|
occurs on the AHA database) may present difficulties for certain two-channel
|
|
analysis methods designed for real-time applications.
|
|
.PP
|
|
Minor tape speed variations should not pose problems for typical arrhythmia
|
|
detectors. It is difficult to avoid tape sticking or slippage during low-speed
|
|
playback, and several episodes of tape slippage were noted and marked with
|
|
comment annotations. Wow and flutter should be studied carefully in the
|
|
context of heart-rate variability studies, since flutter compensation
|
|
was not possible in these recordings. A number of frequency-domain artifacts
|
|
have been identified and related to specific mechanical components of the
|
|
recorders and the playback unit:
|
|
.TS
|
|
center box;
|
|
c | l.
|
|
\fIFrequency (Hz) Source\fR
|
|
_
|
|
0.042 Recorder pressure wheel
|
|
0.083 Playback unit capstan (for twice real-time playback)
|
|
0.090 Recorder capstan
|
|
0.167 Playback unit capstan (for real-time playback)
|
|
0.18\-0.10 Takeup reel (frequency decreases over time)
|
|
0.20\-0.36 Supply reel (frequency increases over time)
|
|
.TE
|
|
The most significant of these artifacts by far is the 0.167 Hz artifact on
|
|
recordings that were played back at real time. The next largest is the
|
|
0.090 Hz artifact; the 0.083 Hz artifact on recordings that were played back
|
|
at twice real-time is of roughly the same magnitude as the 0.090 Hz artifact.
|
|
The 0.042 Hz artifact is of much lower magnitude. Other frequencies
|
|
related to the drive train (at 0.42 Hz, 1.96 Hz, 9.1 Hz, and
|
|
42 Hz) do not appear as noticeable artifacts.
|
|
The frequencies of the last two artifacts listed in the table depend on
|
|
how much tape is on the supply and takeup reels; the supply reel causes
|
|
a much more noticeable artifact than does the takeup reel. Other
|
|
frequency-domain artifacts generated by the supply reel appear in the
|
|
0.10\-0.18 Hz and 0.30\-0.54 Hz bands.
|
|
.PP
|
|
Four of the 48 records (102, 104, 107, and 217) include paced beats.
|
|
The original analog recordings do not represent the pacemaker artifacts
|
|
with sufficient fidelity to permit them to be recognized by pulse amplitude
|
|
(or slew rate) and duration alone, the method commonly used for real-time
|
|
processing. The database records reproduce the analog recordings with
|
|
sufficient fidelity to permit use of pacemaker artifact detectors designed for
|
|
tape analysis, however.
|
|
.SH
|
|
Digitization
|
|
.PP
|
|
The analog outputs of the playback unit
|
|
were filtered to limit analog-to-digital converter (ADC)
|
|
saturation and for anti-aliasing,
|
|
using a passband from 0.1 to 100 Hz relative to real time, well
|
|
beyond the lowest and highest frequencies recoverable from the recordings.
|
|
The bandpass-filtered signals were
|
|
digitized at 360 Hz per signal relative to real time
|
|
using hardware constructed at the MIT Biomedical Engineering Center and
|
|
at the BIH Biomedical Engineering Laboratory.
|
|
The sampling frequency was chosen to facilitate implementations of 60 Hz
|
|
(mains frequency) digital notch filters in arrhythmia detectors.
|
|
Since the recorders were battery-powered, most of the 60 Hz noise present
|
|
in the database arose during playback.
|
|
In those records that were digitized at twice real time, this noise appears
|
|
at 30 Hz (and multiples of 30 Hz) relative to real time.
|
|
.PP
|
|
Samples were acquired from each signal almost simultaneously (the intersignal
|
|
sampling skew was on the order of a few microseconds). As noted above, analog
|
|
tape skew was several orders of magnitude larger. The ADCs were
|
|
unipolar, with 11-bit resolution over a \(+-5 mV range. Sample values thus
|
|
range from 0 to 2047 inclusive, with a value of 1024 corresponding to zero
|
|
volts.
|
|
.PP
|
|
The 11-bit samples were recorded in 8-bit first difference format (this was
|
|
necessary because of limited mass storage capacity).
|
|
Given the sampling frequency and the resolution of the ADC,
|
|
the difference encoding implies a maximum recordable slew rate of \(+-225 mV/s.
|
|
In practice, this limit was exceeded by the input signals very infrequently,
|
|
only during severe noise on a small number of records.
|
|
The effect on the quality of the recorded signals is totally negligible.
|
|
Since the ECG data files are reconstructed from difference data,
|
|
no information is lost and considerable savings in storage may be
|
|
obtained by converting them back into difference form; a program
|
|
for doing so is included in the package of C-language software available
|
|
from the BIH Biomedical Engineering Laboratory.
|
|
.SH
|
|
Annotations
|
|
.PP
|
|
An initial set of beat labels was produced by a simple slope-sensitive
|
|
QRS detector, which marked each detected event as a normal beat. Two
|
|
identical 150-foot chart recordings were printed for each 30-minute record,
|
|
with these initial beat labels in the margin.
|
|
For each record, the two charts were given to two cardiologists, who worked
|
|
on them independently. The cardiologists added additional beat labels
|
|
where the detector missed beats, deleted false detections as necessary,
|
|
and changed the labels for all abnormal beats. They also added rhythm
|
|
labels, signal quality labels, and comments.
|
|
.PP
|
|
The annotations were transcribed from the paper chart recordings.
|
|
Once both sets of cardiologists' annotations for a given record
|
|
had been transcribed and verified, they were automatically compared
|
|
beat-by-beat, and another chart recording was printed. This chart showed
|
|
the cardiologists' annotations in the margin, with all discrepancies
|
|
highlighted. Each discrepancy was reviewed and resolved by consensus.
|
|
The corrections were transcribed, and the annotations were then analyzed
|
|
by an auditing program, which checked them for consistency and which
|
|
located the ten longest and shortest R-R intervals in each record (to
|
|
identify possible missing or falsely detected beats).
|
|
.PP
|
|
In early copies of the database, most beat labels were placed
|
|
at the R-wave peak, but manually inserted labels were not always
|
|
located precisely at the peak.
|
|
In copies of the database made since 1983, the beat labels have been shifted
|
|
from their original locations.
|
|
The ECG (usually the upper signal) was digitally bandpass-filtered to
|
|
emphasize the QRS complexes, and each beat label was moved to the major local
|
|
extremum, after correction for phase shift in the filter.
|
|
A few noisy beats were manually realigned.
|
|
The result is that annotations generally appear at the R-wave peak, and
|
|
are located with sufficient accuracy to make the reference annotation
|
|
files usable for studies requiring waveform averaging and for
|
|
heart rate variability studies (but note the comments
|
|
with respect to analog tape wow and flutter above).
|
|
In the annotated ECG plots in this book, each label is placed so that the
|
|
fiducial mark for the annotation corresponds to the left edge of the label.
|
|
.PP
|
|
The database contains approximately 109,000 beat labels. Sixteen
|
|
have been corrected in the eight years since the database was first
|
|
released (in records 104, 108, 114, 203, 207, 217, and 222); in addition,
|
|
all of the left bundle branch block beats in record 214 were originally
|
|
labelled as normal beats. The rhythm labels
|
|
have been more substantially revised and now include notations for paced
|
|
rhythm, bigeminy, and trigeminy, which were missing in early copies.
|
|
.ds RH S\s-2YMBOLS\s+2
|
|
.bp
|
|
.ce 1
|
|
\fB\s+2SYMBOLS USED IN ECG PLOTS\s-2\fR
|
|
|
|
.TS
|
|
center box;
|
|
c | lw(4.5i).
|
|
\fISymbol Meaning\fR
|
|
_
|
|
\(bu Normal beat
|
|
L Left bundle branch block beat
|
|
R Right bundle branch block beat
|
|
A Atrial premature beat
|
|
a Aberrated atrial premature beat
|
|
J Nodal (junctional) premature beat
|
|
S Supraventricular premature beat
|
|
V Premature ventricular contraction
|
|
F Fusion of ventricular and normal beat
|
|
[ Start of ventricular flutter/fibrillation
|
|
! Ventricular flutter wave
|
|
] End of ventricular flutter/fibrillation
|
|
e Atrial escape beat
|
|
j Nodal (junctional) escape beat
|
|
E Ventricular escape beat
|
|
P Paced beat
|
|
f Fusion of paced and normal beat
|
|
p Non-conducted P-wave (blocked APB)
|
|
Q Unclassifiable beat
|
|
| Isolated QRS-like artifact
|
|
_
|
|
T{
|
|
Rhythm annotations appear \fIbelow\fP the level used for beat
|
|
annotations:
|
|
T}
|
|
(AB Atrial bigeminy
|
|
(AFIB Atrial fibrillation
|
|
(AFL Atrial flutter
|
|
(B Ventricular bigeminy
|
|
(BII 2\(de heart block
|
|
(IVR Idioventricular rhythm
|
|
(N Normal sinus rhythm
|
|
(NOD Nodal (A-V junctional) rhythm
|
|
(P Paced rhythm
|
|
(PREX Pre-excitation (WPW)
|
|
(SBR Sinus bradycardia
|
|
(SVTA Supraventricular tachyarrhythmia
|
|
(T Ventricular trigeminy
|
|
(VFL Ventricular flutter
|
|
(VT Ventricular tachycardia
|
|
_
|
|
T{
|
|
Signal quality and comment annotations appear \fIabove\fR the level
|
|
used for beat annotations:
|
|
T}
|
|
\fIqq\fR T{
|
|
Signal quality change: the first character (`c' or `n') indicates the quality
|
|
of the upper signal (clean or noisy), and the second character indicates the
|
|
quality of the lower signal
|
|
T}
|
|
U Extreme noise or signal loss in both signals: ECG is unreadable
|
|
M Missed beat (`MISSB' in examples, pp. 99\-177)
|
|
P Pause (`PSE' in examples)
|
|
T Tape slippage (`TS' in examples)
|
|
.TE
|