mirror of
https://github.com/This-weiyuqing/ECG_by_Diffenert_CNN_Test.git
synced 2026-10-18 03:13:39 +08:00
add MIT-BIT
This commit is contained in:
1 parent
fcbea99d01
commit
fdf7d4f840
708 files changed
+11062
-4
No files matched your search
@@ -0,0 +1,336 @@
|
||||
<html>
|
||||
<head>
|
||||
<title>MIT-BIH Arrhythmia Database Directory (Introduction)</title>
|
||||
</head>
|
||||
<body bgcolor="#FFFFFF">
|
||||
|
||||
<a href="mitdbdir.htm"><h1 align=center>MIT-BIH Arrhythmia Database Directory
|
||||
</h1></a>
|
||||
|
||||
<p>
|
||||
<b>Next:</b> <a href="records.htm">Records</a>
|
||||
<b>Up:</b> <a href="mitdbdir.htm#toc">Contents</a>
|
||||
<b>Previous:</b> <a href="foreword.htm">Foreword</a>
|
||||
|
||||
<a name="intro"><h1>Introduction</h1></a>
|
||||
<p>
|
||||
This introduction describes how the database records were
|
||||
obtained, and discusses the characteristics of the recorded signals.
|
||||
Following these notes are annotated ``full disclosure'' plots of the entire
|
||||
database. These can be useful for obtaining an overall
|
||||
impression of the contents of individual records. Following the
|
||||
``full disclosure'' plots are sample ECG strips. These strips
|
||||
were chosen to illustrate the salient features of each record.
|
||||
Next are notes on the important features of each record.
|
||||
These notes also include background information on the subjects,
|
||||
including their medications.
|
||||
At the end of the book are tables of rhythms and annotations, which
|
||||
summarize the contents of the database.
|
||||
These tables can be helpful in finding a record with a specific
|
||||
set of characteristics.
|
||||
|
||||
<a name="selection"><h2>Selection criteria</h2>
|
||||
<p>
|
||||
The source of the ECGs included in the MIT-BIH Arrhythmia Database is a
|
||||
set of over 4000 long-term Holter recordings that were obtained
|
||||
by the Beth Israel Hospital Arrhythmia Laboratory between 1975 and 1979.
|
||||
Approximately 60% of these recordings were obtained from inpatients.
|
||||
The database contains 23 records
|
||||
(numbered from 100 to 124 inclusive with some numbers missing)
|
||||
chosen at random from this set, and 25 records
|
||||
(numbered from 200 to 234 inclusive, again with some numbers missing)
|
||||
selected from the same set to include a variety of rare but clinically
|
||||
important phenomena that would not be well-represented
|
||||
by a small random sample of Holter recordings.
|
||||
Each of the 48 records is slightly over 30 minutes long.
|
||||
<p>
|
||||
The first group is intended to serve as a representative sample of the
|
||||
variety of waveforms and artifact that an arrhythmia detector might
|
||||
encounter in routine clinical use. A table of random numbers was used
|
||||
to select tapes, and then to select half-hour segments of them.
|
||||
Segments selected in this way were excluded only if neither of the two
|
||||
ECG signals was of adequate quality for analysis by human experts.
|
||||
<p>
|
||||
Records in the second group were chosen to
|
||||
include complex ventricular, junctional,
|
||||
and supraventricular arrhythmias and conduction abnormalities. Several
|
||||
of these records were selected because features of the rhythm, QRS
|
||||
morphology variation, or signal quality may be expected to present
|
||||
significant difficulty to arrhythmia detectors; these records have
|
||||
gained considerable notoriety among database users.
|
||||
<p>
|
||||
The subjects were 25 men aged 32 to 89 years, and 22 women aged 23 to 89
|
||||
years. (Records 201 and 202 came from the same male subject.)
|
||||
|
||||
<a name="leads"><h2>ECG lead configuration</h2></a>
|
||||
<p>
|
||||
In most records, the upper signal is a modified limb lead II (MLII),
|
||||
obtained by placing the electrodes on the chest. The lower signal is
|
||||
usually a modified lead V1 (occasionally V2 or V5, and in one instance V4);
|
||||
as for the upper signal, the electrodes are also placed on the chest.
|
||||
This configuration is routinely used by the BIH Arrhythmia Laboratory.
|
||||
Normal QRS complexes are usually prominent in the upper signal.
|
||||
The lead axis for the lower signal may be nearly orthogonal to the mean
|
||||
cardiac electrical axis, however (i.e., normal beats are usually
|
||||
biphasic and may be nearly isoelectric).
|
||||
Thus normal beats are frequently difficult to discern in the lower signal,
|
||||
although ectopic beats will often be more prominent (see, for example, record
|
||||
106).
|
||||
A notable exception is record 114, for which the signals were reversed.
|
||||
Since this happens occasionally in clinical practice, arrhythmia detectors
|
||||
should be equipped to deal with this situation.
|
||||
In records 102 and 104, it was not possible to use modified lead II because of
|
||||
surgical dressings on the patients;
|
||||
modified lead V5 was used for the upper signal in these records.
|
||||
|
||||
<a name="analog"><h2>Analog recording and playback</h2></a>
|
||||
<p>
|
||||
The original analog recordings were made using nine Del Mar Avionics
|
||||
model 445 two-channel recorders, designated <i>A</i> through <i>I</i>:
|
||||
<table border>
|
||||
<tr><th><i>Recorder</i></th><th><i>Records</i></th></tr>
|
||||
<tr><td align=center><i>A</i></td><td>102, 107, 111, 115, 121</td></tr>
|
||||
<tr><td align=center><i>B</i></td><td>212</td></tr>
|
||||
<tr><td align=center><i>C</i></td><td>203</td></tr>
|
||||
<tr><td align=center><i>D</i></td><td>118, 124, 217</td></tr>
|
||||
<tr><td align=center><i>E</i></td>
|
||||
<td>101, 103, 106, 108, 112, 117, 119, 122, 209, 219, 220, 223, 233</td></tr>
|
||||
<tr><td align=center><i>F</i></td>
|
||||
<td>104, 109, 123, 205, 207, 210, 215, 221</td></tr>
|
||||
<tr><td align=center><i>G</i></td>
|
||||
<td>100, 105, 114, 116, 213, 214, 222, 228</td></tr>
|
||||
<tr><td align=center><i>H</i></td><td>113, 201, 202, 231</td></tr>
|
||||
<tr><td align=center><i>I</i></td><td>200, 230, 232, 234</td></tr>
|
||||
</table>
|
||||
<br>
|
||||
(It is not known which recorder was used for record 208.)
|
||||
<p>
|
||||
During the digitization process, the analog recordings were played back
|
||||
on a Del Mar Avionics model 660 unit. The analog tapes used for records
|
||||
112, 115 through 124, 205, 220, 223, and 230 through 234 were played back
|
||||
and digitized at twice real time; the rest were played back at real time
|
||||
using a specially constructed capstan for the model 660 unit.
|
||||
Skew between the two signals was found to be as great as 40
|
||||
milliseconds for some of the analog recorders.
|
||||
In addition to the fixed skew that results from extremely small differences
|
||||
in the orientations of the tape heads on the recorder and the playback unit,
|
||||
microscopic vertical wobbling of the tape, either during recording or playback,
|
||||
introduces a variable skew, which may be comparable in magnitude to the fixed
|
||||
skew.
|
||||
This problem (which also
|
||||
occurs on the AHA database) may present difficulties for certain two-channel
|
||||
analysis methods designed for real-time applications.
|
||||
<p>
|
||||
Minor tape speed variations should not pose problems for typical arrhythmia
|
||||
detectors. It is difficult to avoid tape sticking or slippage during low-speed
|
||||
playback, and several episodes of tape slippage were noted and marked with
|
||||
comment annotations. Wow and flutter should be studied carefully in the
|
||||
context of heart-rate variability studies, since flutter compensation
|
||||
was not possible in these recordings. A number of frequency-domain artifacts
|
||||
have been identified and related to specific mechanical components of the
|
||||
recorders and the playback unit:
|
||||
<table border>
|
||||
<tr><th><i>Frequency (Hz)</i></th><th><i>Source</i></th></tr>
|
||||
<tr><td align=center>0.042</td><td>Recorder pressure wheel</td></tr>
|
||||
<tr><td align=center>0.083</td><td>Playback unit capstan (for twice real-time playback)</td></tr>
|
||||
<tr><td align=center>0.090</td><td>Recorder capstan</td></tr>
|
||||
<tr><td align=center>0.167</td>
|
||||
<td>Playback unit capstan (for real-time playback)</td></tr>
|
||||
<tr><td align=center>0.18-0.10</td>
|
||||
<td>Takeup reel (frequency decreases over time)</td></tr>
|
||||
<tr><td align=center>0.20-0.36</td>
|
||||
<td>Supply reel (frequency increases over time)</td></tr>
|
||||
</table>
|
||||
<br>
|
||||
The most significant of these artifacts by far is the 0.167 Hz artifact on
|
||||
recordings that were played back at real time. The next largest is the
|
||||
0.090 Hz artifact; the 0.083 Hz artifact on recordings that were played back
|
||||
at twice real-time is of roughly the same magnitude as the 0.090 Hz artifact.
|
||||
The 0.042 Hz artifact is of much lower magnitude. Other frequencies
|
||||
related to the drive train (at 0.42 Hz, 1.96 Hz, 9.1 Hz, and
|
||||
42 Hz) do not appear as noticeable artifacts.
|
||||
The frequencies of the last two artifacts listed in the table depend on
|
||||
how much tape is on the supply and takeup reels; the supply reel causes
|
||||
a much more noticeable artifact than does the takeup reel. Other
|
||||
frequency-domain artifacts generated by the supply reel appear in the
|
||||
0.10-0.18 Hz and 0.30-0.54 Hz bands.
|
||||
<p>
|
||||
Four of the 48 records (102, 104, 107, and 217) include paced beats.
|
||||
The original analog recordings do not represent the pacemaker artifacts
|
||||
with sufficient fidelity to permit them to be recognized by pulse amplitude
|
||||
(or slew rate) and duration alone, the method commonly used for real-time
|
||||
processing. The database records reproduce the analog recordings with
|
||||
sufficient fidelity to permit use of pacemaker artifact detectors designed for
|
||||
tape analysis, however.
|
||||
|
||||
<a name="digitization"><h2>Digitization</h2></a>
|
||||
<p>
|
||||
The analog outputs of the playback unit
|
||||
were filtered to limit analog-to-digital converter (ADC)
|
||||
saturation and for anti-aliasing,
|
||||
using a passband from 0.1 to 100 Hz relative to real time, well
|
||||
beyond the lowest and highest frequencies recoverable from the recordings.
|
||||
The bandpass-filtered signals were
|
||||
digitized at 360 Hz per signal relative to real time
|
||||
using hardware constructed at the MIT Biomedical Engineering Center and
|
||||
at the BIH Biomedical Engineering Laboratory.
|
||||
The sampling frequency was chosen to facilitate implementations of 60 Hz
|
||||
(mains frequency) digital notch filters in arrhythmia detectors.
|
||||
Since the recorders were battery-powered, most of the 60 Hz noise present
|
||||
in the database arose during playback.
|
||||
In those records that were digitized at twice real time, this noise appears
|
||||
at 30 Hz (and multiples of 30 Hz) relative to real time.
|
||||
<p>
|
||||
Samples were acquired from each signal almost simultaneously (the intersignal
|
||||
sampling skew was on the order of a few microseconds). As noted above, analog
|
||||
tape skew was several orders of magnitude larger. The ADCs were
|
||||
unipolar, with 11-bit resolution over a ±5 mV range. Sample values thus
|
||||
range from 0 to 2047 inclusive, with a value of 1024 corresponding to zero
|
||||
volts.
|
||||
<p>
|
||||
The 11-bit samples were originally recorded in 8-bit first difference format
|
||||
(this was necessary because of limited mass storage capacity). Given the
|
||||
sampling frequency and the resolution of the ADC, the difference encoding
|
||||
implies a maximum recordable slew rate of ±225 mV/s. In practice, this
|
||||
limit was exceeded by the input signals very infrequently, only during severe
|
||||
noise on a small number of records. The effect on the quality of the recorded
|
||||
signals is totally negligible. On this CD-ROM, the samples have been
|
||||
reconstructed from the first differences and stored as pairs of 12-bit
|
||||
amplitudes packed in triplets of consecutive bytes (for details on the storage
|
||||
format, see <a href="/physiotools/wag/signal-5.htm">signal(5)</a>).
|
||||
|
||||
<a name="annotations"><h2>Annotations</h2></a>
|
||||
<p>
|
||||
An initial set of beat labels was produced by a simple slope-sensitive
|
||||
QRS detector, which marked each detected event as a normal beat. Two
|
||||
identical 150-foot chart recordings were printed for each 30-minute record,
|
||||
with these initial beat labels in the margin.
|
||||
For each record, the two charts were given to two cardiologists, who worked
|
||||
on them independently. The cardiologists added additional beat labels
|
||||
where the detector missed beats, deleted false detections as necessary,
|
||||
and changed the labels for all abnormal beats. They also added rhythm
|
||||
labels, signal quality labels, and comments.
|
||||
<p>
|
||||
The annotations were transcribed from the paper chart recordings.
|
||||
Once both sets of cardiologists' annotations for a given record
|
||||
had been transcribed and verified, they were automatically compared
|
||||
beat-by-beat, and another chart recording was printed. This chart showed
|
||||
the cardiologists' annotations in the margin, with all discrepancies
|
||||
highlighted. Each discrepancy was reviewed and resolved by consensus.
|
||||
The corrections were transcribed, and the annotations were then analyzed
|
||||
by an auditing program, which checked them for consistency and which
|
||||
located the ten longest and shortest R-R intervals in each record (to
|
||||
identify possible missing or falsely detected beats).
|
||||
<p>
|
||||
In early copies of the database, most beat labels were placed
|
||||
at the R-wave peak, but manually inserted labels were not always
|
||||
located precisely at the peak.
|
||||
In copies of the database made since 1983, the beat labels have been shifted
|
||||
from their original locations.
|
||||
The ECG (usually the upper signal) was digitally bandpass-filtered to
|
||||
emphasize the QRS complexes, and each beat label was moved to the major local
|
||||
extremum, after correction for phase shift in the filter.
|
||||
A few noisy beats were manually realigned.
|
||||
This process was applied to all records except record 117 in 1983; the
|
||||
beat labels for record 117 were not realigned until March 1998, however.
|
||||
The result is that annotations generally appear at the R-wave peak, and
|
||||
are located with sufficient accuracy to make the reference annotation
|
||||
files usable for studies requiring waveform averaging and for
|
||||
heart rate variability studies (but note the comments
|
||||
with respect to analog tape wow and flutter above).
|
||||
In the annotated ECG plots produced by <tt>psfd</tt> and <tt>pschart</tt>,
|
||||
and in printed copies of this directory, each label is placed so that the
|
||||
fiducial mark for the annotation corresponds to the left edge of the label.
|
||||
<p>
|
||||
The database contains approximately 109,000 beat labels. Sixteen
|
||||
were corrected in the first seven years after the database was released in 1980
|
||||
(in records 104, 108, 114, 203, 207, 217, and 222); in addition,
|
||||
all of the left bundle branch block beats in record 214 were originally
|
||||
labelled as normal beats. The rhythm labels
|
||||
have been more substantially revised and now include notations for paced
|
||||
rhythm, bigeminy, and trigeminy, which were missing in early copies.
|
||||
<p>
|
||||
In October 1998, a rhythm label in record 203 was corrected. In
|
||||
October 2001, a seventeenth error in the beat labels was discovered
|
||||
and corrected (in record 209). In April 2003, 26 PVC annotations in
|
||||
record 119 were manually realigned by small amounts (up to 74 ms). In
|
||||
May 2003, an eighteenth error in the beat labels was discovered and
|
||||
corrected (in record 214). In April 2005, many of the episodes
|
||||
previously labelled as atrial fibrillation in record 222 were
|
||||
partially or completely relabelled as atrial flutter. In April 2008,
|
||||
three beat labels were corrected (two in record 108, and one in record
|
||||
215). In June 2010, the 22nd and 23rd errors in the beat labels were
|
||||
found and corrected (both in record 203). Thanks to Bob Bruce, Pat
|
||||
Hamilton, Yin Dengfeng, Roger Mark, Sebastian Vasquez, and Mariano
|
||||
Llamedo Soria for finding and reporting these errors.
|
||||
|
||||
<hr>
|
||||
|
||||
<a name="symbols"><h1>Symbols used in plots</h1></a>
|
||||
|
||||
<p>
|
||||
[An expanded and updated version of the table below can be found at
|
||||
<a href="/physiobank/annotations.shtml">
|
||||
<tt>http://www.physionet.org/physiobank/annotations.shtml</tt></a>.]
|
||||
|
||||
<p>
|
||||
<table border>
|
||||
<tr><th><i>Symbol</i></th><th><i>Meaning</i></th></tr>
|
||||
<tr><td><b>·</b> <i>or</i> N</td><td>Normal beat</td></tr>
|
||||
<tr><td>L</td><td>Left bundle branch block beat</td></tr>
|
||||
<tr><td>R</td><td>Right bundle branch block beat</td></tr>
|
||||
<tr><td>A</td><td>Atrial premature beat</td></tr>
|
||||
<tr><td>a</td><td>Aberrated atrial premature beat</td></tr>
|
||||
<tr><td>J</td><td>Nodal (junctional) premature beat</td></tr>
|
||||
<tr><td>S</td><td>Supraventricular premature beat</td></tr>
|
||||
<tr><td>V</td><td>Premature ventricular contraction</td></tr>
|
||||
<tr><td>F</td><td>Fusion of ventricular and normal beat</td></tr>
|
||||
<tr><td>[</td><td>Start of ventricular flutter/fibrillation</td></tr>
|
||||
<tr><td>!</td><td>Ventricular flutter wave</td></tr>
|
||||
<tr><td>]</td><td>End of ventricular flutter/fibrillation</td></tr>
|
||||
<tr><td>e</td><td>Atrial escape beat</td></tr>
|
||||
<tr><td>j</td><td>Nodal (junctional) escape beat</td></tr>
|
||||
<tr><td>E</td><td>Ventricular escape beat</td></tr>
|
||||
<tr><td>/</td><td>Paced beat</td></tr>
|
||||
<tr><td>f</td><td>Fusion of paced and normal beat</td></tr>
|
||||
<tr><td>x</td><td>Non-conducted P-wave (blocked APB)</td></tr>
|
||||
<tr><td>Q</td><td>Unclassifiable beat</td></tr>
|
||||
<tr><td>|</td><td>Isolated QRS-like artifact</td></tr>
|
||||
<tr><td colspan=2 align=center>Rhythm annotations appear <i>below</i> the
|
||||
level used for beat annotations:</td></tr>
|
||||
<tr><td>(AB</td><td>Atrial bigeminy</td></tr>
|
||||
<tr><td>(AFIB</td><td>Atrial fibrillation</td></tr>
|
||||
<tr><td>(AFL</td><td>Atrial flutter</td></tr>
|
||||
<tr><td>(B</td><td>Ventricular bigeminy</td></tr>
|
||||
<tr><td>(BII</td><td>2° heart block</td></tr>
|
||||
<tr><td>(IVR</td><td>Idioventricular rhythm</td></tr>
|
||||
<tr><td>(N</td><td>Normal sinus rhythm</td></tr>
|
||||
<tr><td>(NOD</td><td>Nodal (A-V junctional) rhythm</td></tr>
|
||||
<tr><td>(P</td><td>Paced rhythm</td></tr>
|
||||
<tr><td>(PREX</td><td>Pre-excitation (WPW)</td></tr>
|
||||
<tr><td>(SBR</td><td>Sinus bradycardia</td></tr>
|
||||
<tr><td>(SVTA</td><td>Supraventricular tachyarrhythmia</td></tr>
|
||||
<tr><td>(T</td><td>Ventricular trigeminy</td></tr>
|
||||
<tr><td>(VFL</td><td>Ventricular flutter</td></tr>
|
||||
<tr><td>(VT</td><td>Ventricular tachycardia</td></tr>
|
||||
<tr><td colspan=2 align=center>Signal quality and comment annotations appear <i>above</i>
|
||||
the level used for beat annotations:</td></tr>
|
||||
<tr><td><i>qq</i></td>
|
||||
<td>
|
||||
Signal quality change: the first character (`c' or `n') indicates the quality
|
||||
of the upper signal (clean or noisy), and the second character indicates the
|
||||
quality of the lower signal</td></tr>
|
||||
<tr><td>U</td><td>Extreme noise or signal loss in both signals: ECG is unreadable</td></tr>
|
||||
<tr><td>M (<i>or</i> MISSB)</td><td>Missed beat</td></tr>
|
||||
<tr><td>P (<i>or</i> PSE)</td><td>Pause</td></tr>
|
||||
<tr><td>T (<i>or</i> TS)</td><td>Tape slippage</td></tr>
|
||||
</table>
|
||||
|
||||
<HR>
|
||||
<P><ADDRESS>
|
||||
<I><A HREF="mailto:george@mit.edu">George B. Moody (<tt>george@mit.edu</tt>)</A></ADDRESS></I><BR>
|
||||
24 May 1997
|
||||
<br>
|
||||
<i>Revised 24 June 2010</i>
|
||||
</body>
|
||||
</html>
|
||||
Reference in new issue
Block a user